100% Canadian owned and operated, delivering science Canada wide

Water System, GenPure UV/UF, Barnstead

Product number: 966-155-86
Systems | Barnstead GenPure Water Systems
Ordered upon request
a
product
Brand: Thermo Fisher Scientific
Call for pricing

• Delivers ultrapure 18.2 Mê.cm water with consistent quality.
• All GenPure systems exceed international standards ASTM Type 1, ISO 3696 Grade 1, ASTM D1193 and CLSI-CLRW.
• Flow rate: up to 2L/min.
• Operating pressure min/max: 29-87 psi (2-6 bar).
• Conductivity: 0.055µS/cm.
• Bacteria: <1 CFU/ml. Particles: <1 0.22µm/ml.
• Integrated feed water monitoring: an additional measuring cell monitors feed water conductivity. Any exceeding of the limiting value is immediately displayed maximizing cartridge life.
• USP-compliant conductivity measurement with temperature compensation can be switched on or off.
• Real-time TOC monitoring for continual measurement of organic substances.
• Continuous monitoring of UV intensity removes the risk of false TOC values due to diminished UV intensity.
• Fault message alerts user when the UV lamp intensity is no longer sufficient for accurate TOC measurements.
• UV photo-oxidation 185/254nm: high performance UV assembly reliably reduces the content of microorganisms and their metabolites. UV oxidation also reduces organic compounds to ultra-low levels.
• Tilting control panels for easy viewing.
• Ergonomically designed controller tilts for optimal key pressing and easy readability.
• Conductivity cells are carefully calibrated prior to each measurement via built-in reference resistance with cell constants at 0.01/cm.
• Temperature measurements are made by a platinum chip sensor with ±0.1ºC accuracy.
• Aquastop quick-connect capability enables cartridge replacement within seconds, even mid-operation.
• GLP-compliant documentation includes real time clock and code-protected operating system to prevent unauthorized changes to system settings.
• RS232 interface with adjustable send-interval.
• Digital microprocessor control automatically monitors and stores faults from the last 4 weeks.
• All systems include a wall bracket, ultrapure polisher cartridge, sterile 0.2µm filter, UV lamp and/or ultrafilter where applicable.
• Dimensions (W x D x H): 14.6 x 13 x 24.2”.
• Power: 90 to 240V. CSA approved.

Feed water requirements:
• Source: potable tap water, pretreated by reverse osmosis, ion exchange or distillation.
• Feed water conductivity: <2µS/cm.
• TOC: max 50 ppb.
• Bacteria count: <100 CFU/ml.
• Turbidity: <1.0 NTU.
• Temperature: 2 to 35ºC.
• Pressure: 1.4 to 87 psi (0.1 to 6 bar).
• Feed water connection: 0.75” NPT.