Haze Turbidity Meter for Beer

Product number: 960-845-02
Portable | Haze Turbidity Meter for Beer
Ordered upon request
a
product
productproduct
Brand: Hanna Instruments
Order in - 960-845-02
Please login to place an order

This Haze Meter is specially designed for measurement of haze in beer quality analysis. The instrument is based on a state-of-the-art optical system which guarantees accurate results, assures long term stability, and minimizes stray light and color interferences. It also compensates for variations in intensity of the lamp, although proper calibration ensures measurements are fully validated, comparable, and in compliance with regulatory requirements. The 25 mm round cuvettes composed of special optical glass guarantee the repeatability of turbidity measurements.

ASBC Compliant Measurement (American Society of Brewing Chemists) - Measurements are taken from beer samples that have been chilled and degassed, and results are reported in Formazin Turbidity Units (FTU).

Calibration –A two, three, or four-point turbidity calibration can be performed by using the supplied (<0.1, 15, 100, and 800 FTU) standards. Calibration points can be modified if user-prepared standards are used.

AMCO AEPA-1 Primary Turbidity Standard - These non-toxic standards are made of styrene divinylbenzene polymer spheres that are uniform in size and density. The standards are reusable and stable with a long shelf life.

Fast Tracker - Allows brewers to record the time and location of a specific measurement or series of measurements using iButton tags on specific tanks for quick and easy readings. 

Turbidity Specifications
Range: 0.00 to 1000 FTU
Resolution: 0.01 (0.00 to 9.99 FTU); 0.1 (10.0 to 99.9 FTU); 1 (100 to 1000 FTU)
Range Selection: automatic
Accuracy: ±2% of reading plus 0.05 FTU
Repeatability: ±1% of reading or 0.02 FTU, whichever is greater
Stray Light: < 0.1 FTU
Light Detector: silicon photocell
Method: ratio nephelometric method
Measuring Mode: normal, average, continuous
Turbidity Standards: < 0.1, 15, 100, and 800 NTU
Calibration: two, three, or four-point calibration

General Specifications
Light Source: LED @ 580 nm
Display: 60 x 90 mm backlit LCD
Log Memory: 200 records
Connectivity: USB
Environment: 0 to 50°C (32 to 122°F), RH max 95% non-condensing
Power Supply: 1.5V AA alkaline batteries (4) or AC adapter; auto-off after 15 minutes of non-use
Dimensions: 224 x 87 x 77 mm (8.8 x 3.4 x 3.0”)
Weight: 512 g (18 oz.)
Supplied with sample cuvettes and caps (6), HI847492-11 calibration cuvette, 25 mL glass vials with caps (4), cuvette cleaning cloth, batteries, AC adapter, instrument quality certificate, instruction manual, and rugged carrying case.

Warranty: 2 years