No need to worry about DO probe compatibility – the meter accepts and automatically recognizes Orion polarographic and RDO optical/luminescence-based probes.
Dissolved oxygen: (concentration – polarographic) 0.00 to 90.0mg/L ±0.2. (% saturation) 0.0 to 600.0 ±2%. (concentration – RDO) 0.00 to 50.00mg/L ±0.1mg/L up to 8mg/L; ±0.2mg/L from 8 to 20mg/L; 10% of reading from 20 to 50mg/L.
Dissolved oxygen: (% saturation – RDO) 0.0 to 500.0 ±2%. Manual barometric pressure correction: 450.0 to 850.0mm Hg. Manual salinity factor correction: 0 to 45ppt.
Temperature range: 0.0 to 50.0ºC (32 to 122ºF) ±0.1. Offset calibration: 1 point.
Log Function: Manual or automatic with AUTO-READ™. Log edit: delete last reading or all.
Waterproof with a IP67-rated housing.
Large, backlit graphic LCD display.
Display shows dissolved oxygen in % saturation or concentration with temperature and electrode status, time, date, sample ID, user ID and calibration points.
Advanced features such as stability and averaging options provide additional options allowing for laboratory accuracy everywhere you need it.
Accurate results are easy with ATC, manual salinity and automatic/manual barometric pressure compensation.
Calibrate using water-saturated air, air-saturated water, a Winkler titration or zero-point calibration.
AUTO-READ™ locks in the stable reading on your screen, ready indicator alerts when readings are stable and timed reading gathers data in specific time intervals.
Non-volatile memory holds up to 5000 data points with time and date stamp.
Easily transfer data and keep meter software up-to-date with the USB and RS232 ports and complimentary data analysis software.
Multilingual interface supports English, Spanish, French, Italian, German and Chinese.
Inputs: DO with temperature probe = 9-pin mini-DIN.
Output: RS232 and USB.
Power: 4 x AA batteries provide over 800 hours of operation. Can also use the optional AC adapter (120V).