The Haier Biomedical UltraEco series of upright -86°C ultra-low temperature freezers stores biological samples, vaccines, reagents and other sensitive materials long term, with some of the lowest energy use in its class.
How to Choose
- 730 L (25.8 cu. ft.): holds up to 500 two-inch boxes (20 racks); 41.2 in. wide.
- 830 L (29.3 cu. ft.): holds up to 600 two-inch boxes (24 racks); 45.1 in. wide.
- Check your space: both are 77.6 in. tall and 39.3 in. deep. Allow room around the unit for airflow and to open the door fully.
- Power: both run on 100–230 V, 50/60 Hz.
Reading the Comparison Table
- Capacity: usable storage volume in litres and cubic feet
- Cryobox Capacity (2"): how many standard 2-inch sample boxes fit, and in how many racks
- Exterior / Interior Dimensions: width x depth x height of the cabinet and of the storage space
- Power Consumption: average daily energy use, in kWh/day
- Max. Power: peak electrical draw, for sizing the circuit
- Noise Level: operating sound level; lower is quieter
Key Features
- Energy efficient: variable-speed compressor and natural hydrocarbon (HC) refrigerants with ultra-low global-warming potential cut operating costs by up to 50% versus the previous generation. ENERGY STAR® certified.
- Precise temperature control: adjustable from -40°C to -86°C in 1°C steps, with dual sensors and fast recovery after door openings.
- 10-inch touchscreen: sample management; shows cabinet temperature, setpoint, ambient temperature and input voltage. An LED light bar shows operating status at a glance.
- Alarms: high/low temperature, high condenser temperature, power failure, high/low voltage, sensor error, low battery, high ambient temperature and door ajar, plus compressor self-diagnostics.
- Organized storage: 4 insulated inner doors and 3 adjustable stainless steel shelves.
- Connectivity: USB, RS-232/485 and remote alarm contacts; two 15 mm rear access ports.
- Options: CO2 or LN2 backup system and temperature recorder.
Certifications: CE, UL, ENERGY STAR.