Typical applications include soil and agricultural analysis, biological fluids, petrochemical industry, and food and drink analysis.
Uses a soft touch key pad for ease of data entry and features electronic auto-ignition, automatic flame failure safety and will operate using butane, natural gas or LPG. Backlit, alphanumeric LCD display.
Three modes: general analytical applications, serum and urine.
Range (General analytical applications mode) Na: 0.1 to 100ppm, up to 250meq/L, 1:100 dil; K: 0.1 to 100ppm, up to 250meq/L, 1:100 dil; Ca: 15 to 100ppm, up to 250meq/L, 1:100 dil; Li: 0.5 to 100ppm, up to 250meq/L, 1:100 dil.
Range (Serum) Na: 0 to 200, 1:100 dil; K: 0 to 10, 1:100 dil; Ca: 0 to 2, 1:5 dil; Li: 0 to 2, 1:5 dil.
Range (Urine) Na: 0 to 250, 1:100 dil; K: 0 to 100, 1:100 dil; Ca: 0 to 10, 1:2 dil; Li: 0 to 10, 1:2 dil.
Resolution: 0.1ppm/meq.
Data displayed in concentration units (ex. Ppm, meq, etc.).
Fully adjustable gas control.
Axial flow type atomiser.
Silicon photodiode detector.
Up to 5-point calibration with curve fitting software. Calibration curve fit routines first order and second order. Multiple calibration curve saving facility.
Sample data storage.
Printer attachment.
Flame system: LPG and dry oil free air.
RS232 interface.
Compressor air supply: oil-free compressor unit with pressure regulator.
Combustion gas LPG is controlled by precision regulator.
Supplied with air compressor with buffer tank, gas ignitor, 5cc beakers and filters for sodium, potassium, lithium and calcium.
Dimensions (L x W x H): (main unit) 365 x 245 x 220mm; (compressor) 290 x 255 x 210mm.